Korekcja BTPS |
automatyczna |
Spirometria wdechowa/wydechowa |
tak/tak |
Spirometria natężona płuc |
FVC, Best FVC, FEV0.75, FEV1, Best FEV1, FEV3, FEV6, PEF, FEV0.75/FVC, FEV1/, FVC,
FEV3/FVC, FEV6/FVC, FEV0.75/SVC, FEV1/SVC, FEV3/SVC, FEV6/SVC, PIF, FIVC, FIV1,
MEF75, MEF50, MEF25, FEF75, FEF50, FEF25, MMEF, FET25, FET50, MIF75, MIF50,
MIF25, PEFT, FIF50, FEF50/FIF50, FEF50/SVC, FEV0.75/ FEV6, FEV1/ FEV6, FIV1/FIVC,
VEXT, wiek płuc
BTL CardioPoint Spiro: Aex, FEV075/VC, FEV1/VC, FEV3/VC, FEV6/VC, FEF25-75,
FEF50/VC, VEXT, T0/PEF, T200/PEF, VEXT/FVC |
Testy bronchoprowokacyjne PRE/POST |
tak |
Spirometria swobodna |
SVC, ERV, IRV, TV, IC, IVC |
Maksymalna wentylacja dowolna MVV |
MVV, MVVf, MRf |
Wartości wzorcowe |
ECCS 1983, ECCS/ERS 1993, Zapletal 1977, Roca, Barcelona 1986, NHANES III 1999,
Knudson 1983, Knudson 1976, ITS 1984, Crapo 1981, Lam 1982, Pereira 1996, Gore 1995,
GLI 2012, Hou Shu 1990, Jia Ju-cai 1990, Sun Bin 1990, Liu Shi-Wan 1990, Liu Guo-Hua 1990,
Zhu Xi 1990, Wu 1961, Ip 2006, Polgar 1979, Wang Yang 2013, Perez Padilla 2003, Platino 2006,
PDPI 1992, Thai 2000 |
Automatyczna interpretacja |
ATS, BTS, Enright, GOLD |
System motywacyjny dla dzieci (animacja) |
tak |
Zakres przepływu (litry/s) |
16 l/s (wdech/wydech) |
Dokładność (50 ml/s do 16 l/s) |
± 5% lub 50 ml (ważna większa wartość) |
Zakres objętości (litry) |
0.025 do 8 litrów |
Dokładność (0.025 do 8 l) |
± 3% lub 50 ml (ważna większa wartość) |
Opór przepływu |
< 79 Pa / l/s |
Czujnik temperatury |
+ 10°C do + 40°C |
Dokładność pomiaru temperatury |
± 2% |
Czujnik wilgotności względnej |
0 do 100% RH |
Dokładność pomiaru wilgotności |
± 4% dla 25°C, 30%–80% RH |
Rozdzielczość ADC |
15 bitów |
Częstotliwość próbkowania |
1000 Hz |